GB/T 14710-2009 'Environmental Requirements and Test Methods for Medical Electrical Appliances' was proposed by the National Food and Drug Administration, updated and released on November 15, 2009, and officially implemented on May 1, 2010. The purpose of the standard is to evaluate the adaptability of medical electrical equipment in various working environments and simulated storage and transportation environments by applying certain environmental stresses.
Is it necessary to refer to the GB/T 14710-2009 standard when environmental testing is required for medical electrical equipment?
The answer from the Medical Device Technical Evaluation Center of the National Drug Administration is as follows:
If the mandatory standards applicable to the products to be declared comply with the GB/T 14710-2009 standard, testing should be carried out in accordance with the GB/T 14710-2009 standard and the relevant requirements of the mandatory standard.
Interpretation: If the mandatory standards applicable to the product have clear guidelines, guiding principles, etc., as long as these documents have clear requirements to comply with the GB/T 14710 standard, they must be implemented according to the requirements;
If the product to be declared is used in special environments (such as high temperature, high humidity, or low temperature), supporting information that the product can be used in the corresponding environment should be provided in the research materials.
Interpretation: If all mandatory standards (including recommended standards and guiding principles) applicable to the product do not require the application of the GB/T 14710 standard, then it is also necessary to consider whether the registered and declared product will be used in special environments (such as high temperature, low temperature, high humidity, or vibration). If it is confirmed that it will be used, then it is still necessary to provide supporting information that the product can be used in that environment, The test report passed the environmental test verification in accordance with GB/T 14710-2009.
Standard Overview:
This standard specifies the purpose, environmental grouping, transportation testing, adaptability to power supply, reference test conditions, special circumstances, test procedures, test sequence, test requirements, test methods, and detailed rules that should be specified when referencing this standard for environmental testing of medical electrical equipment.
Applicable scope of the standard:
Applicable to all electrical equipment or systems that meet the definition of medical devices.
The following content will summarize the content of the GB/T 14710-2009 standard and summarize the environmental testing requirements and methods for medical electrical equipment:
2. Environmental grouping of medical electrical equipment:
The equipment is divided into three groups based on climatic and mechanical environments, as follows:
2.1 Climate and environmental grouping (according to usage conditions):
a) Group I
Use in a good environment. Usually refers to the use of equipment in a controlled environment with air conditioning and other equipment;
b) Group II
Used in general environments. Usually refers to the use of equipment in an environment with optimal supply and ventilation;
c) Group III
Used in harsh environments. Usually refers to the use of equipment in non insulated and ventilated environments, as well as similar outdoor environments.
2.2 Mechanical environment grouping (based on transportation and circulation conditions):
a) Group I
Equipment that is carefully operated and subjected to slight vibrations and impacts during transportation and circulation generally refers to fixed and rarely moved equipment;
b) Group II
Equipment that is allowed to be subjected to general vibration and impact during use, generally referring to equipment that is easy to move;
c) Group III
Equipment that is subject to vibration and impact during frequent transportation, loading and unloading, and handling.
3. Environmental condition grouping:
3. 成品包装运输试验要求
在设备正常出厂包装条件下,设备应按标志“向上'的位置捆在载重汽车的后部,试验时汽车的负荷量应为额定载重量的1/3。
行车路面:按JTG B01-2003标准规定的三级公路。
行车距离:200 km。
行车速度:30 km/h~40 km/ho
按上述要求,试验可以使用载重汽车进行,也可以使用运输试验装置进行。试验完毕,检验设备紧固件是否有松动现象并按产品标准规定的检验项目进行检测。
4. 对电源的适应能力
4. 1由电网电源供电的设备,试验电压为额定值的110%或90%。
4.2对电源频率及电源电压有特殊要求的设备,其频率、电压的工作范围、试验方法可在产品标准中另行规定。
5. 基准试验条件
基准试验条件基准值或范围,及允差见表2 。
在不产生疑义时,可在+10℃〜+40℃,相对湿度30%〜75%,交流电压为:额定值±10%,电源频率为:额定值±2 %的条件下进行。
3. Explanation of special circumstances
6.1 If the test conditions specified by the manufacturer for the test group are not completely consistent with those specified in Table 1, the manufacturer's regulations should be followed.
6.2 When individual influence quantities cannot be tested according to the benchmark test conditions specified in this standard, the test should be conducted according to the manufacturer's specified conditions.
When it is not feasible to conduct a complete machine test, it is allowed to divide the equipment into several parts for testing, and the manufacturer should specify which key components or parts to test.
6.4 If the equipment (such as large equipment or equipment with specific requirements for the operating environment) is not suitable for certain environmental testing items (such as rated working low temperature test, rated working high temperature test, rated working damp heat test, etc.), the manufacturer should provide specific testing requirements and explain them in the product standard.
When the equipment is applicable to other national or industry standards, and these standards also specify environmental testing requirements, environmental testing should be carried out according to their regulations.
4. Test Procedure
Each test typically includes the following procedures:
a) Pre processing (if necessary);
b) Initial testing (if necessary);
c) Testing;
d) Intermediate testing (if necessary);
e) Operational tests (if necessary);
f) Restore (if necessary);
g) Final inspection.
5. Test sequence
When conducting multiple tests on the same equipment in sequence, the tests are generally carried out in the following order:
a) Rated working low temperature test;
b) Low temperature storage test;
c) Rated working high temperature test;
d) High temperature storage test;
e) Rated working damp heat test;
f) Damp heat storage test;
g) Vibration test;
h) Collision test;
i) Transportation test.
If the test sequence is affected, it shall be specified by the product standard.